Alert burden: interactive demo

A model with a good AUC can still produce more alerts than a ward can act on, because real events are rare and every patient is scored many times. In this demo an early-warning model with an AUC of 0.90, used at a threshold with 90% sensitivity, sends a nurse with 6 patients about 22 alerts in a 12-hour shift, one every 33 minutes, and about 1 alert in 199 is for a real event. The threshold should be chosen from that workload as well as from sensitivity.

In one line: AUC 0.90 at 90% sensitivity, 6 patients scored hourly: 22 alerts per nurse per 12-hour shift, one every 33 minutes, about 1 in 199 real.

SYNTHETIC DATA

These demos are illustrative. Every signal, image, patient and note is synthetic and fictional, generated and scored in your browser, and every number shown is computed from that synthetic data. They show general principles, not the performance of any real model, product or project, and none of them is a medical device.

Nothing is uploaded. No patient data is used.

14

Clinical model validation / workload

A good AUC. An alert every half hour.

What this shows: a threshold is also a workload. Counted per nurse per shift, a model with a strong AUC can raise an alert every few minutes, almost all of them false, because real events are rare and every patient is scored many times.

Set the model and the ward

A shift is 12 hours. A real event, such as deterioration, happens in 2% of patient-shifts and is present at one of that patient’s scores.

One nurse, seven shifts

each mark is an alert
Alerts over seven shiftsAlerts for one nurse over seven shifts.

Amber: false alertGreen: alert for a real event

Alerts per nurse per shift—
Real alerts—
Time between alerts—
Real events missed—

Synthetic data · illustrative · not a medical device

Where this matters

Early-warning scores, sepsis and deterioration alerts, and any model whose output interrupts a clinician. Express the threshold as alerts per clinician per shift and real events per alert, and agree what load the ward can act on before choosing it.

What this experiment models

Scores are normally distributed for assessments with and without an event, separated by the amount that gives the chosen AUC. The threshold is the one that reaches the chosen sensitivity. Each patient is scored at regular intervals through a 12-hour shift; 2% of patient-shifts contain one real event, present at one of those scores. Alerts per shift is the number of scores times the share that exceed the threshold. The chart simulates every score for one nurse over seven shifts with a fixed seed. Repeated alerts on the same patient are counted separately, as they are when each one has to be acknowledged.

The principle

How it works

An early-warning model scores each patient repeatedly and raises an alert when the score crosses a threshold. The threshold is usually chosen to reach a target sensitivity, and the report gives the AUC and that sensitivity. Neither number says how often the alert sounds. That depends on three more things: how rare real events are, how many patients a clinician looks after and how often each one is scored.

In the demo a real event happens in 2% of patient-shifts. With six patients scored every hour, a nurse’s shift contains 72 scores, almost all of them for patients who are fine. Even a small false-positive rate on 72 scores gives many alerts, while real events arrive about once every nine shifts. The chart shows one simulated week for one nurse, with each alert as a mark.

When nearly every alert is false, alerts stop being acted on, which is the problem usually called alarm fatigue. The remedies are a better model, a higher threshold with a stated cost in missed events, less frequent scoring, or suppressing repeats for the same patient. Each is a decision for the clinical team, and it can only be made when the threshold is expressed as workload. See clinical model validation.

Worked example

An early-warning model with an AUC of 0.90

Picture a hospital introducing a model that warns of patient deterioration. The validation report shows an AUC of 0.90. The clinical lead asks for a threshold that catches 90% of events, which sounds like a careful choice, and the model is switched on for a ward where each nurse looks after six patients.

The demo’s default settings are that ward: AUC 0.90, 90% sensitivity, 6 patients per nurse, each scored every hour. Alerts per nurse per shift reads 22, and Time between alerts reads 33 min. Real alerts reads 1 in 199. Real events missed reads 10%, which is what a sensitivity of 90% means.

The chart shows a simulated week for one nurse: 151 alerts, of which 0 are real. That is not bad luck. With a real event in 2% of patient-shifts, a nurse with six patients meets one about once every 9 shifts, and in between every alert is false.

Now look at what would change it. A model with an AUC of 0.95 gives 11 alerts a shift, 1 in 100 real. At 0.98 it is 3.9 a shift, 1 in 36. Keeping the 0.90 model and lowering the sensitivity to 80% gives 12 alerts a shift, 1 in 125 real, at the price of missing 20% of events. Scoring once a shift instead of hourly gives 1.9 alerts, 1 in 17 real.

The correct practice is to choose the threshold with the people who will receive the alerts. Express it as alerts per clinician per shift and real events per alert for the actual staffing and scoring frequency. Agree what load can be acted on and what miss rate is acceptable. Consider suppressing repeat alerts for a patient who has already been reviewed, and measure the alert rate and the response to alerts after go-live.

Checklist

What to check in your own project

  • How many alerts per clinician per shift does the chosen threshold produce on your ward?
  • How many alerts are raised for each real event?
  • How often is each patient scored, and are repeat alerts for the same patient suppressed?
  • What share of real events is missed at this threshold, and is that accepted by the clinical team?
  • What is a clinician expected to do for each alert, and how long does it take?
  • Are the alert rate and the response to alerts measured after deployment?

Related: our clinical model validation service, the wiki page on medical AI validation, and the demos on calibration and decision thresholds and predictive value and prevalence.

FAQ

Common questions

How many alerts will a clinical prediction model generate?

Multiply the number of scores per clinician per shift by the share that exceed the threshold. That share depends on the model’s AUC, the sensitivity you ask for and how rare events are. In this demo an AUC of 0.90 at 90% sensitivity, with 6 patients scored hourly, gives about 22 alerts per nurse per 12-hour shift.

Why does a model with a high AUC cause alarm fatigue?

Because AUC says nothing about how rare the event is or how often patients are scored. When events are rare, even a modest false-positive rate applied to many scores produces far more false alerts than real ones. In the demo about 1 alert in 199 is for a real event.

How do I choose a threshold for an early-warning score?

Start from the workload the ward can act on and the miss rate it can accept, and find the threshold that fits both. In the demo, lowering sensitivity from 90% to 80% cuts alerts from 22 to 12 a shift and raises missed events from 10% to 20%. That trade-off is a clinical decision and should be made explicitly.

How can alert burden be reduced without missing more events?

Improve discrimination, score less often, or suppress repeat alerts for a patient who has already been reviewed. In the demo, raising the AUC from 0.90 to 0.95 at the same 90% sensitivity halves the alerts, from 22 to 11 a shift. Scoring once a shift instead of hourly cuts them to 1.9, although a real system would then detect events later.

What is a reasonable number of alerts per nurse per shift?

There is no universal figure; it depends on what each alert requires and on everything else the nurse is doing. What matters is that the number is known before deployment and agreed with the ward. The demo reports it alongside the time between alerts and the number of alerts per real event.

Medical AI lab

More demos

01 / Heart rate from ECGHeart rate from a noisy ECG – interactive demoMissed and false beats behind a correct average02 / Patient leakagePatient-level data leakage – interactive demoRecord-level vs patient-level train/test splits03 / Calibration & thresholdsCalibration and decision thresholds – interactive demoCalibration plot, decision curve and net benefit04 / Scanner shiftScanner shift in medical imaging – interactive demoSegmentation tuned on one scanner, tested on another05 / Dice vs clinical errorDice score vs clinical error – interactive demoDice, HD95 and lesions detected06 / Note de-identificationDe-identifying a clinical note – interactive demoEnglish, Hebrew and mixed notes vs a rule-based detector07 / Nodule volume 3DNodule volume and CT slice thickness – 3D interactive demoThe same nodule, measured on thin and thick slices08 / Sample sizeSample size for validating a clinical AI model – interactive demoHow wide the interval is around a measured sensitivity09 / SubgroupsHidden subgroup failure in clinical AI – interactive demoA good overall score next to a weak subgroup10 / Slice-by-slice 3DSlice-by-slice segmentation of a 3D lesion – 3D interactive demoA high per-slice Dice and a lesion reported in pieces11 / Predictive value 3DPositive predictive value and prevalence – 3D interactive demoWhat a positive result means when the condition is rare12 / Shortcut learningShortcut learning in medical imaging AI – interactive demoA model that reads the hospital tag, not the disease13 / DriftModel drift and performance monitoring – interactive demoHow late a monitoring rule notices a slow decline

MLAIA DATA SCIENCE

Tell us what
you’re working on.

Share the problem, the data you have and what success would look like. We’ll discuss a practical next step.

Please don’t include confidential datasets, credentials or patient information.

yochai@mlaia.com
WhatsApp: +972 52 484 6282

Your inquiry is handled under our Privacy Policy.